Sientra Primary and Revision Augmentation Rupture Trending and Analysis With Magnetic Resonance Imaging
Author Type(s)
Faculty
DOI
10.1093/asj/sjv021
Journal Title
Aesthetic Surgery Journal
First Page
33
Last Page
42
Document Type
Article
Publication Date
5-1-2015
Department
Medicine
Keywords
Adolescent, Adult, Aged, Breast, Breast Implantation, Breast Implants, Female, Follow-Up Studies, Humans, Magnetic Resonance Imaging, Microscopy, Electron, Scanning, Middle Aged, Prospective Studies, Prosthesis Design, Reoperation, Rupture, Silicone Gels, Therapeutic Irrigation, Young Adult
Disciplines
Medicine and Health Sciences
Abstract
BACKGROUND: Rupture of silicone gel breast implants is a rare occurrence but remains one of the key surgical concerns. The objective of this article was to provide visibility and information on trends for the impact that patient and surgical characteristics play in the occurrence of rupture.
OBJECTIVES: Examine trends in surgical techniques to better understand the etiology of implant rupture.
METHODS: Analysis was based on Sientra's prospective, open-label, U.S.-based clinical study of High-Strength Cohesive silicone breast implants. Patient and surgical characteristics were compared between ruptured and intact implants.
RESULTS: The subset of data used for this analysis included 1792 implants in 935 primary and revision augmentation patients implanted by 31 plastic surgeons, with an average follow-up of 6.6 years. The results confirm that rupture remains a rare adverse event. Overall, the rupture prevalence for this study was 2.4%. Rupture prevalence was lower among textured devices (0.8%) compared to smooth devices (3.8%). The prevalence of rupture was 7.8% among devices placed with a transaxillary incision site compared to 1.6% and 3.0% when placed with an inframammary or periareolar incision site, respectively. Rupture was reported in 5.5% of the devices that received steroid pocket irrigation, compared to 1.8% of the devices that did not.
CONCLUSIONS: Although ruptures in the Sientra study with the High-Strength Cohesive silicone gel implants were an uncommon occurrence, the authors were able to identify strong trends for the association of certain surgical factors and characteristics. The results show among other factors that an inframammary approach and a textured device were found to be protective against rupture.
LEVEL OF EVIDENCE: 2 Therapeutic.
Recommended Citation
Haws, M., Kaufman, D. L., & Alizadeh, K. (2015). Sientra Primary and Revision Augmentation Rupture Trending and Analysis With Magnetic Resonance Imaging. Aesthetic Surgery Journal, 35 (Suppl. 1), 33-42. https://doi.org/10.1093/asj/sjv021